Remote Control Commands
R&S
®
SMW200A
1067User Manual 1175.6632.02 ─ 16
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Configuring the LF generator" on page 1062.
Manual operation: See "State" on page 511
[:SOURce]:LFOutput<ch>:VOLTage <Voltage>
Sets the output voltage of the selected LF output.
Parameters:
<Voltage> float
Range: dynamic (see data sheet)
Increment: 0.001
*RST: 1
Default unit: V
Example:
SOURce:LFOutput1:VOLTage 1.5
Manual operation: See "Output Voltage" on page 511
[:SOURce]:LFOutput<ch>:INTernal:VOLTage <Voltage>
Sets the output voltage for the LF generators.
The sum of both values must not exceed the overall output voltage, set with command
[:SOURce]:LFOutput<ch>:VOLTage.
Suffix:
<ch>
.
[1]|2
LF1 and LF2
Parameters:
<Voltage> float
Range: 0 to 4
*RST: 0.5
Example: See Example "Configuring the LF generator" on page 1062.
[:SOURce<hw>]:LFOutput<ch>:SHAPe <Shape>
Selects the waveform shape of the LF signal.
Parameters:
<Shape> SINE | SQUare | TRIangle | TRAPeze
*RST: SINE
Example: See Example "Configuring the LF generator" on page 1062.
Options: TRIangle|TRAPeze require R&S SMW-K24
Manual operation: See "Shape" on page 504
SOURce Subsystem