Remote Control Commands
R&S
®
SMW200A
906User Manual 1175.6632.02 ─ 16
Parameters:
<Source> INTB | INTernal | OBASeband | EGT1 | EGT2 | EGC1 | EGC2 |
ELTRigger | INTA | ELCLock | BEXTernal | EXTernal | BBSY
*RST: INTernal
Example: See Example "Adjusting clock, marker and trigger settings"
on page 892
Options: ELTRigger|ELCLock require R&S SMW-B10
BBSY require R&S SMW-B9
Manual operation: See "Trigger Source" on page 256
[:SOURce<hw>]:BB:DM:TRIGger:SLENgth
<SLength>
Defines the length of the signal sequence to be output in the SINGle trigger mode.
Parameters:
<SLength> integer
Range: 1 to 4294967295
*RST: 1000
Default unit: symbol
Example: See Example "Adjusting clock, marker and trigger settings"
on page 892
Manual operation: See "Trigger Signal Duration" on page 255
[:SOURce<hw>]:BB:DM:TRIGger:RMODe?
Queries the status of signal generation.
Return values:
<RMode> STOP | RUN
Example:
SOURce1:BB:DM:TRIGger:SOURce ELTRigger
SOURce1:BB:DM:TRIGger:SEQuence ARETrigger
SOURce1:BB:DM:TRIGger:RMODe?
Response: RUN
Usage: Query only
Manual operation: See "Running/Stopped" on page 255
[:SOURce<hw>]:BB:DM:TRIGger:EXTernal:SYNChronize:OUTPut <Output>
Enables signal output synchronous to the trigger event.
Parameters:
<Output> 0 | 1 | OFF | ON
*RST: 1
Example: See Example "Adjusting clock, marker and trigger settings"
on page 892
SOURce Subsystem