Remote Control Commands
R&S
®
SMW200A
908User Manual 1175.6632.02 ─ 16
Example: See Example "Specifying delay and inhibit values in time units"
on page 922
Usage: Query only
Manual operation: See "Actual Trigger Delay/Actual External Delay" on page 258
[:SOURce<hw>]:BB:DM:TRIGger:OBASeband:TDELay <ObasTimeDelay>
Specifies the trigger delay for triggering by the signal from the other path.
Parameters:
<ObasTimeDelay> float
Range: 0 to 7929.170398682
Increment: 250E-12
*RST: 0
Default unit: s
Example: See Example "Specifying delay and inhibit values in time units"
on page 922
Manual operation: See "(Specified) External Delay/(Specified) Trigger Delay"
on page 257
[:SOURce<hw>]:BB:DM:TRIGger:OBASeband:INHibit <Inhibit>
Specifies the number of symbols by which a restart is inhibited. This command applies
only for triggering by the second path.
Parameters:
<Inhibit> integer
Range: 0 to 67108863
*RST: 0
Default unit: symbol
Example: See Example "Adjusting clock, marker and trigger settings"
on page 892
Manual operation: See "External / Trigger Inhibit" on page 257
[:SOURce<hw>]:BB:DM:TRIGger:DELay:UNIT <TrigDelUnit>
Determines the units the trigger delay is expressed in.
Parameters:
<TrigDelUnit> SAMPle | TIME
*RST: SAMPle
Example: See Example "Specifying delay and inhibit values in time units"
on page 922
Manual operation: See "(External) Delay Unit" on page 257
SOURce Subsystem