Remote Control Commands
R&S
®
SMW200A
1029User Manual 1175.6632.02 ─ 16
[:SOURce<hw>]:CORRection:CSET:DATA:POWer <Power>
Enters the level values to the table selected with [:SOURce<hw>]:CORRection:
CSET[:SELect].
Parameters:
<Power> Power#1[, Power#2, ...]
String of values with default unit dB.
Example: See Example "Create a table with user-defined correction values
for the RF level" on page 1026 .
Manual operation: See "Edit List Mode Data" on page 487
[:SOURce<hw>]:CORRection:CSET:DATA:FREQuency:POINts?
[:SOURce<hw>]:CORRection:CSET:DATA:POWer:POINts?
Queries the number of frequency/level values in the selected table.
Return values:
<Points> integer
Range: 0 to 10000
*RST: 0
Example: See Example "Create a table with user-defined correction values
for the RF level" on page 1026 .
Usage: Query only
[:SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe
Fills the selected user correction table with the level values measured by the power
sensor for the given frequencies.
To select the used power sensor set the suffix in key word SENSe.
Example:
See Example "Fill user correction data with sensor"
on page 1027.
Usage: Event
Manual operation: See " Fill User Correction Data with Sensor" on page 537
[:SOURce<hw>]:CORRection:CSET[:SELect] <Filename>
Selects or creates a file for the user correction data.
If the file with the selected name does not exist, a new file is created.
Parameters:
<Filename> string
Filename or complete file path; file extension can be omitted.
SOURce Subsystem