EasyManua.ls Logo

Rohde & Schwarz SMW200A - Page 902

Rohde & Schwarz SMW200A
1312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMW200A
887User Manual 1175.6632.02 ─ 16
Table 13-3: Value ranges of the suffixes ENTity<ch> and SOURce<hw> in advanced configuration
with multiple entities
SCPI Syntax
ENTity<ch> SOURce<hw>
SOURce<hw>:BB:...
- 1 .. 8
ENTity<ch>:SOURce<hw>:BB:...
1 .. 8 1 .. 4
See also Chapter 13.3, "SCPI Command Aliases for Advanced Mode with Multiple
Entities", on page 767.
The following commands are available:
[:SOURce<hw>]:BBIN:FOFFset.......................................................................................887
[:SOURce<hw>]:BB:FOFFset..........................................................................................887
[:SOURce<hw>]:BB:POFFset..........................................................................................887
[:SOURce<hw>]:BB:PGAin............................................................................................. 888
[:SOURce<hw>]:BBIN:PGAin.......................................................................................... 888
[:SOURce<hw>]:BB:ROUTe............................................................................................888
[:SOURce<hw>]:BBIN:ROUTe.........................................................................................888
[:SOURce<hw>]:BB:CODer:MODE..................................................................................888
[:SOURce<hw>]:BB:POWer:PEAK?.................................................................................889
[:SOURce<hw>]:BB:CFACtor?........................................................................................ 889
[:SOURce<hw>]:BB:POWer:RMS?.................................................................................. 889
[:SOURce<hw>]:BBIN:FOFFset <FOffset>
[:SOURce<hw>]:BB:FOFFset <FOffset>
Sets a frequency offset for the internal/external baseband signal. The offset affects the
generated baseband signal.
Parameters:
<FOffset> float
Range: depends on the installed options, e.g. -60 MHz to
+60 MHz (R&S SMW-B10)
Increment: 0.01
*RST: 0
Default unit: Hz
Example:
BB:FOFF 2MHZ
Sets a frequency offset of 2 MHz.
Options: BBIN:FOFFset requires R&S SMW-B10
Manual operation: See "Frequency Offset" on page 386
[:SOURce<hw>]:BB:POFFset <POffset>
Sets the relative phase offset for the selected baseband signal compared to the base-
band signals of the other baseband sources (second path or external baseband).
SOURce Subsystem

Table of Contents

Related product manuals