Remote Control Commands
R&S
®
SMW200A
1148User Manual 1175.6632.02 ─ 16
Setting parameters:
<Source> AUTO | IMMediate | SINGle | BUS | EXTernal | EAUTo
AUTO [IMMediate]
Executes a sweep automatically.
In this free-running mode, the trigger condition is met continu-
ously. I.e. when a sweep is completed, the next one starts
immediately.
SINGle [BUS]
Executes one complete sweep cycle.
The following commands initiate a trigger event:
*TRG on page 773
[:SOURce<hw>]:SWEep:POWer:EXECute
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute
:TRIGger<hw>[:SWEep][:IMMediate], :TRIGger<hw>:
PSWeep[:IMMediate] and :TRIGger<hw>:FSWeep[:
IMMediate].
Set the sweep mode with the commands:
[:SOURce<hw>]:SWEep:POWer:MODE AUTO|STEP
[:SOURce<hw>]:SWEep[:FREQuency]:MODE AUTO|STEP
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:MODE
AUTO|STEP
In step mode (STEP), the instrument executes only one step.
EXTernal
An external signal triggers the sweep.
EAUTo
An external signal triggers the sweep. When one sweep is fin-
ished, the next sweep starts.
A second trigger event stops the sweep at the current frequency,
a third trigger event starts the trigger at the start frequency, and
so on.
*RST:
AUTO
Example: See Example "Setup an LF sweep" on page 1061
Usage: Setting only
Manual operation: See "Mode" on page 476
:TRIGger<hw>:FSWeep[:IMMediate]
:TRIGger<hw>:PSWeep[:IMMediate]
:TRIGger<hw>:LFFSweep:IMMediate
:TRIGger<hw>[:SWEep][:IMMediate]
Performs a single sweep and immediately starts the activated, corresponding sweep:
●
FSWeep - RF frequency
●
PSWeep - RF level
●
LFFSweep - LF frequency
●
SWEep - all sweeps
TRIGger Subsystem