Remote Control Commands
R&S
®
SMW200A
852User Manual 1175.6632.02 ─ 16
Usage: Event
Manual operation: See "Apply and Connect" on page 158
:SCONfiguration:EXTernal:CODer<ch>:REMote:INFO?
:SCONfiguration:EXTernal:FADer<ch>:REMote:INFO?
:SCONfiguration:EXTernal:BBMM<ch>:REMote:INFO?
:SCONfiguration:EXTernal:IQOutput<ch>:REMote:INFO?
Queries information on the external instrument.
Return values:
<Info> string
Returns information in one of the following formats:
"None"
An external instrument is not assigned
<SymbolicName>,<RemoteChannel>,<RemoteAd-
dress>,<RF Path>
An external instrument is assigned, but not connected
<SymbolicName>,<RemoteChannel>,<RemoteAd-
dress>,<RF Path>[,<*IDN? String>]
An external instrument is assigned and connected
Example:
:SCONfiguration:EXTernal:BBMM1:REMote:INFO?
"SGT (101676)","LAN","rssgt100a101676","A",
"Rohde&Schwarz,SGT100A,1419.4501k02/101676,3.1.19.4-3.18.251.99"
Usage: Query only
Manual operation: See "Remote Connection Status" on page 152
:SCONfiguration:EXTernal:CODer<ch>:REMote:DISConnect
:SCONfiguration:EXTernal:FADer<ch>:REMote:DISConnect
:SCONfiguration:EXTernal:BBMM<ch>:REMote:DISConnect
:SCONfiguration:EXTernal:IQOutput<ch>:REMote:DISConnect
Disconnects the selected remote connection. To disconnect all remote connections at
once, use the command :SCONfiguration:EXTernal:REMote:DISConnect[:
ALL].
Example:
See Example "Connecting and configuring external instrument
(advanced configuration)" on page 826
Usage: Event
Manual operation: See "Remote Connection Status" on page 152
:SCONfiguration:EXTernal:REMote:INITialization:PREDefined:CATalog?
:SCONfiguration:EXTernal:REMote:INITialization:USER:CATalog?
Queries the names of the existing user defined/predefined initialization files.
SCONfiguration Subsystem