Remote Control Commands
R&S
®
SMW200A
882User Manual 1175.6632.02 ─ 16
SOURce:PGENerator:OUTPut:STATe 1
SOURce:PULM:STATe 1
OUTPut1:STATe 1
The following commands are available:
[:SOURce<hw>]:PULM:SOURce..................................................................................... 882
[:SOURce<hw>]:PULM:PERiod....................................................................................... 882
[:SOURce<hw>]:PULM:DELay........................................................................................ 883
[:SOURce<hw>]:PULM:DOUBle:DELay............................................................................883
[:SOURce<hw>]:PULM:DOUBle:WIDTh........................................................................... 883
[:SOURce<hw>]:PULM:POLarity..................................................................................... 883
[:SOURce<hw>]:PULM:IMPedance..................................................................................884
[:SOURce<hw>]:PULM:MODE........................................................................................ 884
[:SOURce<hw>]:PULM:OUTPut:VIDeo:POLarity...............................................................884
[:SOURce]:PULM[:INTernal][:TRAin]:TRIGger:IMMediate...................................................885
[:SOURce<hw>]:PULM:STATe........................................................................................885
[:SOURce<hw>]:PULM:TRIGger:MODE...........................................................................885
[:SOURce<hw>]:PULM:TRIGger:EXTernal:IMPedance......................................................885
[:SOURce<hw>]:PULM:TTYPe........................................................................................885
[:SOURce<hw>]:PULM:WIDTh........................................................................................886
[:SOURce<hw>]:PULM:DOUBle:STATe........................................................................... 886
[:SOURce<hw>]:PULM:SOURce <Source>
Selects between the internal (pulse generator) or an external pulse signal for the mod-
ulation.
Parameters:
<Source> INTernal | EXTernal
*RST: INTernal
Example: See Example "Perform pulse modulation" on page 881.
Manual operation: See "Source" on page 513
[:SOURce<hw>]:PULM:PERiod <Period>
Sets the period of the generated pulse, that means the repetition frequency of the inter-
nally generated modulation signal.
Parameters:
<Period> float
The minimum value depends on the installed options
R&S SMW-K22 or R&S SMW-K23
Range: 20E-9 to 100
Increment: 5E-9
*RST: 10E-6
Example: See Example "Perform pulse modulation" on page 881.
Manual operation: See "Pulse Period" on page 515