Remote Control Commands
R&S
®
SMW200A
1039User Manual 1175.6632.02 ─ 16
// ******************************************************************
// If the attenuation between the interconnection LO OUT and LO IN is
// greater than 1 dB at the oscillator frequency,
// we recommend that you perform the LO level adjustment
// ******************************************************************
:CALibration:LOSCillator:Coupling:LOCAL?
// ******************************************************************
// Address and reset the third instrument
// ******************************************************************
*RST; *CLS
// ******************************************************************
// Configure the oscillator signal of the third instrument
// ******************************************************************
SOURce:FREQuency 5 GHz
SOURce:FREQuency:LOSCillator:MODe EXTernal
SOURce:FREQuency:LOSCillator:OUTPut:STATe OFF
// ******************************************************************
// If the attenuation between the interconnection LO OUT and LO IN is
// greater than 1 dB at the oscillator frequency,
// we recommend that you perform the LO level adjustment
// ******************************************************************
:CALibration:LOSCillator:Coupling:LOCAL?
See also:
●
"To configure LO coupling with cascaded instruments to achieve phase coherence"
on page 454
●
"Level adjustment" on page 449
[:SOURce<hw>]:FREQuency:LOSCillator:INPut:FREQuency?......................................... 1040
[:SOURce<hw>]:FREQuency:LOSCillator:MODE............................................................ 1040
[:SOURce<hw>]:FREQuency:LOSCillator:OUTPut:FREQuency?......................................1041
[:SOURce<hw>]:FREQuency:LOSCillator:OUTPut:STATe............................................... 1041
[:SOURce<hw>]:FREQuency:MODE..............................................................................1041
[:SOURce<hw>]:FREQuency[:CW|FIXed].......................................................................1042
[:SOURce<hw>]:FREQuency[:CW|FIXed]:RCL............................................................... 1043
[:SOURce<hw>]:FREQuency:MANual............................................................................1043
[:SOURce<hw>]:FREQuency:MULTiplier........................................................................1044
[:SOURce<hw>]:FREQuency:OFFSet............................................................................ 1044
[:SOURce<hw>]:FREQuency:CENTer............................................................................1045
[:SOURce<hw>]:FREQuency:SPAN...............................................................................1045
[:SOURce<hw>]:FREQuency:STARt..............................................................................1045
[:SOURce<hw>]:FREQuency:STOP...............................................................................1046
[:SOURce<hw>]:FREQuency:STEP:MODE.................................................................... 1046
[:SOURce<hw>]:FREQuency:STEP[:INCRement]........................................................... 1047
SOURce Subsystem