6. BASIC OPERATION
97
• Operation keys Other Than S-CUT
Other operation keys can be controlled from the mouse and touch panel.
This is valid on the video signal waveform display and vector display. Each time you press
the key, the settings assigned to the key switch.
[See also] Video signal waveform display → 10.1, “Operation Key Description”
Vector display → 11.1, “Operation Key Description”
6.5 Customized Layout (SER26)
The layout of the measurement screen that appears when the WFM, VECT, PIC, AUDIO,
STATUS, or EYE key is pressed (one type each) and the screen that appears when the MULTI
key is pressed (six types) can be arranged freely. To do so, connect a mouse to the front panel
USB port or use the touch panel.
The layout that you specify will not be initialized even if you initialize the instrument. To
initialize, perform any of the procedures below.
[See also] 7.7, “Initialization”
LAYOUT INITIALIZE YES on the SYS menu: The entire layout is initialized.
ALL INITIALIZE YES on the SYS menu: The entire layout is initialized.
Factory default settings: The entire layout is initialized.
DEFAULT LAYOUT in the layout window The layout of the selected measurement
display is initialized.
6.5.1 Notes
The maximum number of items that can be arranged is determined by 128 ÷ the number of
display channels (1 to 4).
Item, here, includes the items on the Main and Sub tabs as well as Format, Input, and Time
on the Option tab.