Command reference
R&S
®
ZNB/ZNBT
1057User Manual 1173.9163.02 ─ 62
Parameters:
<Result> S | Y | Z
S-parameters, Y-parameters, Z-parameters
Example:
*RST; CALC:PAR:MEAS 'Trc1'", '"Y-S22'
Select the converted admittance Y <-- S22 as measurement
parameter of the default trace.
CALC:TRAN:COMP S
Convert the converted Y-parameter into an S-parameter.
CALCulate<Chn>:TRANsform:IMPedance:RNORmal <Model>
Selects the theory for the renormalization of port impedances. The selection has an
impact on the conversion formulas for wave quantities and S-parameters.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Model> TWAVes | PWAVes
TWAVes - travelling waves
PWAVes - power waves
*RST: TWAVes
Example: See [SENSe<Ch>:]PORT<PhyPt>:ZREFerence
Manual operation: See "Renormalization According to Theory of" on page 294
7.3.1.14 CALCulate:TTIMe:...
Defines the properties and retrieves the results of the rise time measurement provided
with the Extended time domain analysis option R&S ZNB/ZNBT-K20.
CALCulate<Chn>:TTIMe:DATA?....................................................................................1057
CALCulate<Chn>:TTIMe:STATe.................................................................................... 1058
CALCulate<Chn>:TTIMe:THReshold............................................................................. 1058
CALCulate<Chn>:TTIMe:DATA? [<Data>]
Queries the results of the rise time measurement
Suffix:
<Chn>
.
Channel number used to identify the active trace
Query parameters:
<Data> ALL
If omitted, a single numeric value is returned. If ALL is specified,
the result consists of 6 numeric values. Furthermore, the inter-
pretation of the result values depends on the current trace's
stimulus axis (CALCulate<Chn>:TRANsform:TIME:XAXis).
See the table below.
SCPI command reference