GUI reference
R&S
®
ZNB/ZNBT
508User Manual 1173.9163.02 ─ 62
Remote command:
The port parameters in many calibration commands define ports to be calibrated.
Type
Selects the calibration type to be performed. For background information, see Chap-
ter 4.5.7.2, "Calibration types", on page 200.
SMARTerCal is based on a "One Path Two Ports" or a full n-port calibration (n > 1):
"Adapter Removal" is a 2-port calibration, all other calibration types can be used for
any combination of 2 or more ports.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
Power Meter
Shows a list of all power meters that have been properly configured and are available
for the scalar source power calibration of a single source port. The last configured
power meter is selected by default. See "Configured Devices" on page 782.
Remote command:
[SENSe:]CORRection:COLLect:PMETer:ID
Step 2: Connectors and Cal Kits
Selects the connector type and gender for all ports and allows you to import a calibra-
tion kit.
The upper part of the panel shows the previously configured SMARTerCal(s). Select
the appropriate one to get access to the related port settings. For the individual calibra-
tions, the functionality is the same as described for the single calibration version of the
wizard (see "Step 2: Connectors and Cal Kits" on page 500).
Back
Go back to Step 2: Connectors and Cal Kits.
Cal softtool