GUI reference
R&S
®
ZNB/ZNBT
447User Manual 1173.9163.02 ─ 62
Start Auto Cal
Performs a fully automatic system error calibration for the active channel, using the
active calibration unit with its factory calibration and auto-detection of ports.
If one port is active, an OSM calibration is performed. If two ports are active, an UOSM
calibration is performed.
Calibration – Start... (Cal Unit)
Depending on the Multiple Cal in Calibration Wizard setting, the "Start... (Cal Unit)" but-
ton either opens the Calibration Unit wizard or the Calibration Unit wizard (MultiCal).
Calibration – Start... (Manual)
Depending on the Multiple Cal in Calibration Wizard setting, the "Start... (Cal Unit)" but-
ton either opens the Calibration Setting wizard or the Calibration Setting wizard (Multi-
Cal).
Calibration – Repeat...
Re-enters the data acquisition step of the current calibration, restoring the related
channel settings. This provides a convenient way to repeat or correct an existing cali-
bration without necessarily repeating all measurements.
Scalar Power Cal – Power Cal...
Opens the Power Cal wizard.
If the active recall set contains several channels, an "Info" box lets you decide whether
the R&S ZNB/ZNBT acquires calibration data for the active channel or for all channels.
SMARTerCal – Start... (Cal Unit)
Depending on the Multiple Cal in Calibration Wizard setting, the "Start... (Cal Unit)" but-
ton either opens the SMARTerCal wizard (Cal Unit) or the SMARTerCal wizard (Cal
Unit, MultiCal).
SMARTerCal – Start... (Manual)
Depending on the Multiple Cal in Calibration Wizard setting, the "Start... (Cal Unit)" but-
ton either opens the SMARTerCal wizard (manual) or the SMARTerCal wizard (manual,
MultiCal).
SMARTerCal – Repeat...
Reenters the data acquisition step of the current SMARTerCal, restoring the related
channel settings. This function provides a convenient way to repeat or correct an exist-
ing SMARTerCal without necessarily repeating all measurements.
Cal Validation ...
Opens the Cal Validation dock widget that allows you to set up and run a Calibration
validation.
5.11.1.2 Channel Selection for Calibration dialog
If you initiate a calibration and there is more than one channel in the current recall set,
you are prompted to select the channels to be calibrated:
Cal softtool

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals