GUI reference
R&S
®
ZNB/ZNBT
626User Manual 1173.9163.02 ─ 62
Note: The loaded file is stored in the active recall set. Recall sets contain the full
(de-)embedding data so that they can be transferred to other instruments.
Remote command:
MMEMory:LOAD:VNETworks<Ch>:SENDed:DEEMbedding<PhyPt>
MMEMory:LOAD:VNETworks<Ch>:SENDed:EMBedding<PhyPt>
Fixture Tool
The "Fixture Tool" section is only available for deembedding 2-port data networks.
Its controls allow you to select and run a third-party fixture modeling tool (see Chap-
ter 4.6.2.9, "Fixture modeling and deembedding", on page 217) and to use its results
for single-ended deembedding.
If the selected fixture modeling tool is not installed on the instrument, the "Run Tool"
button is disabled/grayed out. Otherwise it opens the Fixture modeling dialog that
guides you through the fixture modeling.
Use the "Info" button to get additional information about the selected tool.
Remote command:
CALCulate:FMODel:DEASsistant:TOOL
5.13.5 Fixture modeling dialog
Allows you to model a test fixture using the selected fixture modeling tool and to deem-
bed selected ports using the generated touchstone files.
Supported fixture modeling tools
The following fixture modeling tools are supported:
Chapter 4.7.8, "Eazy de-embedding based on IEEE 370", on page 250
Chapter 4.7.9, "In-situ de-embedding", on page 251
Chapter 4.7.10, "Smart fixture de-embedding", on page 252
Access: Channel – [Offset Embed] > "Single Ended" / "Balanced" > "Run Tool"
Offset Embed softtool