Command reference
R&S
®
ZNB/ZNBT
1348User Manual 1173.9163.02 ─ 62
SOURce:GROup:SIMultaneous:FOFFset:MOFFSet:
DVALue 0 MHz
Set minimum frequency offset for frequency shifted parallel mea-
surement to 0 MHz.
:SOURce:GROup:SIMultaneous:FOFFset:MOFFset:MODE
BANDwidth
:SOURce:GROup:SIMultaneous:FOFFset:MOFFset:
BWFactor 100
Set the minimum frequency offset for frequency shifted parallel
measurement to at least 100 times the measurement bandwidth
Manual operation: See "Parallel Measurement with Frequency Offset" on page 576
7.3.15.3 SOURce:POWer:CORRection:COLLect...
The SOURce:POWer:CORRection:COLLect... commands allow you to set up a
scalar source power calibration.
SOURce:POWer:CORRection:COLLect:AVERage[:COUNt] <NoReadings>
Sets a limit for the number of calibration sweeps in the source power calibration. The
command is valid for all channels, ports and external generators.
Equivalent command: SOURce:POWer:CORRection:NREadings.
Parameters:
<NoReadings> Number of readings
Range: 1 to 100
*RST: 2
Example: See SOURce<Ch>:POWer:CORRection[:ACQuire] (for
equivalent command SOURce:POWer:CORRection:
NREadings)
Manual operation: See "Flatness Cal – Max Iterations" on page 525
SOURce:POWer:CORRection:COLLect:AVERage:NTOLerance <Tolerance>
Specifies the maximum deviation of the measured power from the target power of the
calibration. The command is valid for all channels and calibrated ports.
Parameters:
<Tolerance> Tolerance value
Range: 0.001 dB to 1000 dB
*RST: 1 dB
Default unit: dB
Example: See SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation: See "Flatness Cal – Tolerance" on page 525