EasyManua.ls Logo

R&S ZNB Series - 7.3.1.11 CALCulate:RIPPle...

R&S ZNB Series
1611 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNB/ZNBT
976User Manual 1173.9163.02 ─ 62
Similar to CALCulate<Ch>:PARameter:SDEFine, but in presence of balanced ports
this command distinguishes between
single-ended S-parameters for logical ports (SSS<Li><Lj>)
"raw" single-ended S-parameters referring to physical ports (S<Pi><Pj>)
Suffix:
<Ch>
.
Channel number
Setting parameters:
<TraceName> Trace name
<Result> Measured quantity
Usage: Setting only
Manual operation: See "S-Parameter (selector)" on page 285
CALCulate<Ch>:PARameter:SELect <TraceName>
Selects an existing trace as the active trace of the channel. All trace commands with-
out explicit reference to the trace name act on the active trace (e.g.
CALCulate<Chn>:FORMat). CALCulate<Ch>:PARameter:SELect is also neces-
sary if the active trace of a channel has been deleted.
Suffix:
<Ch>
.
Channel number.
Parameters:
<TraceName> Trace name, e.g. 'Trc4'. See "Rules for trace names" in "Table
Area" on page 337.
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named Ch4Tr1 to measure the
input reflection coefficient S
11
. The trace is the active trace in
channel 4.
CALC4:PAR:SDEF 'Ch4Tr2', 'S22'
Create another trace named Ch4Tr2 to measure the output
reflection coefficient S
22
. Again this new trace becomes the
active trace in channel 4.
CALC4:PAR:SEL 'Ch4Tr1'
Select the first trace Ch4Tr1 as the active trace.
CALC4:FORM MLIN
Calculate the magnitude of S
11
and display it in a linearly scaled
Cartesian diagram.
7.3.1.11 CALCulate:RIPPle...
The CALCulate:RIPPle... commands define the ripple limits and control the ripple
limit check.
SCPI command reference

Table of Contents