GUI reference
R&S
®
ZNB/ZNBT
365User Manual 1173.9163.02 ─ 62
Infinite Average/Restart
"Infinite Average" enables or disables "Infinite Averaging " (default: disabled), "Restart"
starts a new averaging cycle.
Remote command:
CALCulate<Chn>:IAVerage[:STATe]
Mode
Selects the quantities to be averaged.
"Magnitude -
Phase"
Perform averaging of magnitude and phase (default) of the complex
trace value.
Real - Imag
Perform averaging of the real and imaginary part of the complex trace
value (i.e. over the complex value itself).
Remote command:
CALCulate<Chn>:IAVerage:MODE
5.5.11 Trace Data tab
Stores one or several data or memory traces to a file or loads a memory trace from a
file.
Background information
Refer to Chapter 4.4.2, "Trace files", on page 151.
All buttons on the "Trace Data" tab serve as "openers" for related dialogs:
"Import..." calls up a dialog to load a memory trace from a trace file; see Chap-
ter 5.5.11.1, "Import Complex Data dialog", on page 366.