Page #159 background imageLoading...
Page #159 background image
Concepts and features
R&S
®
ZNB/ZNBT
159User Manual 1173.9163.02 ─ 62
Renormalization of S-parameters
Renormalization means that the S-parameters at connector impedances are converted
to S-parameters at certain target impedances.
During Touchstone export, the S-parameters can be renormalized in two ways:
Common target impedance
The <reference resistance> of the Option line is taken as common target
impedance and the data is renormalized to the common target impedance, regard-
less of the reference impedances of the involved ports.
If the impedances of all involved ports are identical, the analyzer sets the
<reference resistance> in the Option line to the real part of this impedance.
Otherwise it sets it to 50 (Ω).
Port-specific Reference impedances
In this case, the reference impedances of the individual ports are used for the
renormalization and these impedances are documented in the Renormalization
information comment. If the port-specific reference impedances are different, an
additional warning is added to this comment, indicating that the option line contains
a non-matching reference resistance.
Note that when reimporting this type of file into standard applications (including the
R&S ZNB/ZNBT itself), the <reference resistance> of the Touchstone option
line is used and the impedance system underlying the data is not interpreted cor-
rectly.
4.4.2.2 ASCII (*.csv) files
An ASCII file contains a header and the actual trace data:
freq;reTrc1_S21;imTrc1_S21;reMem2[Trc1]_S21;imMem2[Trc1]_S21;
300000.000000;0.000000;0.000000;0.000000;0.000000;
40499497.487437;0.000000;0.000000;0.000000;0.000000;
80698994.974874;0.494927;-0.065174;0.500833;-0.074866;
120898492.462312;0.497959;-0.111724;0.488029;-0.107375;
...
The header consists of the following data elements:
<Stimulus> stimulus variable: freq for Frequency sweep, power for Power sweep,
time for Time sweep, trigger for CW Mode sweep.
<reTrace1> first response value of first trace: re<Trace_Name>,
mag<Trace_Name> or db<Trace_Name> for output format Re/Im, lin. Mag-Phase
or dB Mag-Phase, respectively. The data format for export files can be selected in
the Export Data dialog.
<imTrace1> second response value of first trace: im<Trace_Name> for output for-
mat Re/Im, ang<Trace_Name> for output formats lin. Mag-Phase or dB Mag-
Phase. The data format for export files can be selected in the Export Data dialog.
<reTrace2> first response value of second trace: re<Trace_Name>,
mag<Trace_Name> or db<Trace_Name> for output format Re/Im, lin. Mag-Phase
Operations on traces

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals