6.5.4 Application of the status reporting system
The purpose of the status reporting system is to monitor the status of one or several
devices in a measuring system. To do this and react appropriately, the controller must
receive and evaluate the information of all devices. The following standard methods
described in the following sections are used:
Service request (SRQ) initiated by the measuring device
Serial poll of all devices in the bus system, initiated by the controller to find out who
sent an SRQ and why
Parallel poll of all devices
Query of a specific instrument status by commands
Query of the error queue
6.5.4.1 Service request
The R&S ZNB/ZNBT can send a service request (SRQ) to the controller. Usually this
service request causes an interrupt, to which the control program can react appropri-
ately.
Initiating an SRQ
As shown in section Overview of status registers, an SRQ is initiated if one or several
of bits 2, 3, 4, 5 or 7 of the status byte are set and enabled in the SRE. Each of these
bits summarizes the information of a further register, the error queue or the output buf-
fer.
The ENABle parts of the status registers can be set such that arbitrary bits in an arbi-
trary status register initiate an SRQ. To use the possibilities of the service request
effectively, all bits in the enable registers SRE and ESE must be set to "1".
Example: Use *OPC to generate an SRQ
1. Set bit 0 in the ESE (Operation Complete).
2. Set bit 5 in the SRE (ESB).
3. Insert *OPC in the command sequence (e.g. at the end of a sweep).
When all commands preceding *OPC have been completed, the instrument generates
an SRQ.
Example: Generate an SRQ when a limit is exceeded
1. Set bit 3 in the SRE (summary bit of the STATus:QUEStionable register, set
after STATus:PRESet)
2. Set bit 10 in the STATus:QUEStionable:ENABle register (summary bit of the
STATus:QUEStionable:LIMit1 register)
3. Set bit 1 in the STATus:QUEStionable:LIMit1:ENABle register
Status reporting system

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals