Command reference
R&S
®
ZNB/ZNBT
846User Manual 1173.9163.02 ─ 62
CALCulate:CALValidate:STANdard <Standard>
Selects the CalU standard to be used for cal validation.
Parameters:
<Standard> OPEN | SHORt | MATCh
7.3.1.2 CALCulate:DATA...
The CALCulate:DATA... commands provide access to the results of a measure-
ment.
Data format
The trace data is transferred in either ASCII or block data (REAL) format, depending
on the FORMat[:DATA] setting. If block data format is used, it is recommended to
select EOI as a receive terminator (SYSTem:COMMunicate:GPIB[:SELF]:
RTERminator EOI).
CALCulate<Chn>:DATA..................................................................................................846
CALCulate:DATA:ALL?................................................................................................... 849
CALCulate<Ch>:DATA:CALL...........................................................................................850
CALCulate<Ch>:DATA:CALL:CATalog?............................................................................851
CALCulate<Ch>:DATA:CHANnel:ALL?.............................................................................851
CALCulate<Ch>:DATA:CHANnel:DALL?.......................................................................... 851
CALCulate:DATA:DALL?.................................................................................................852
CALCulate<Ch>:DATA:MDATa:INTerpolate.......................................................................852
CALCulate<Chn>:DATA:NSWeep:COUNt?.......................................................................852
CALCulate<Chn>:DATA:NSWeep:FIRSt?......................................................................... 852
CALCulate<Chn>:DATA:NSWeep[:LAST]?....................................................................... 853
CALCulate<Ch>:DATA:SGRoup?.................................................................................... 854
CALCulate<Chn>:DATA:STIMulus?................................................................................. 854
CALCulate:DATA:TRACe................................................................................................ 855
CALCulate<Chn>:DATA <Format>, <Data>...
CALCulate<Chn>:DATA? <Format>
The query reads the response values of the selected channel's active trace or reads
error terms of the selected channel.
The "set command" either imports formatted or unformatted trace data to the selected
channel's active trace (see Chapter 4.1.5, "Data flow", on page 98) or writes error
terms of the selected channel.
Note
The data format is parameter-dependent; see tables below. The unit is the default
unit of the measured parameter; see CALCulate<Ch>:PARameter:SDEFine.
Unformatted trace data (SDATa) can only be imported to memory traces
Formatted trace data (FDATa) can only be imported to "live" traces if the related
channel is in single sweep mode (INITiate<Ch>:CONTinuous OFF).
SCPI command reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals