849User Manual 1173.9163.02 ─ 62
Error Term Description Receive Ports (S-parameter)
SCORr14 Source match 3 (S33)
SCORr15 Reflection tracking 3 (S33)
SCORr16 Isolation 3 (S31)
SCORr17 Load match 3 (S31)
SCORr18 Transmission tracking 3 (S13)
SCORr19 Isolation 1 (S13)
SCORr20 Load match 1 (S13)
SCORr21 Transmission tracking 1 (S13)
SCORr22 Isolation 3 (S32)
SCORr23 Load match 3 (S32)
SCORr24 Transmission tracking 3 (S32)
SCORr25 Isolation 2 (S23)
SCORr26 Load match 2 (S23)
SCORr27 Transmission tracking 2 (S23)
Note: The error terms are channel-specific; they apply to the active calibration of chan-
nel no. <Chn> or to the factory calibration (if no channel calibration is active). For the
factory calibration, the query form is allowed only (no change of factory calibration
data).
Tip: Use the generalized command [SENSe<Ch>:]CORRection:CDATa to read or
write error terms for arbitrary analyzer ports. For additional programming examples
refer to Chapter 8.2.5.3, "Saving and recalling error terms", on page 1518.
CALCulate:DATA:ALL? <Format>[, <RecallSet>]
Reads the current response values of all traces of the referenced recall set.
Query parameters:
<Format> FDATa | SDATa | MDATa
Output format for the S-parameter data, see
CALCulate<Chn>:DATA.
<RecallSet> Recall set name; if omitted the active recall set is used
Return values:
<Data> Response values either in ASCII or block data format, depend-
ing on the current FORMat[:DATA] setting.
Example: See CALCulate<Chn>:DATA
Usage: Query only