Example:
RST; SOUR:POW:CORR:TCO:DEF 1GHz, -5, DUT; DEF
2GHz, -10, DUT; DEF2?
Define two points in the power loss list of the two-port in front of
the DUT; query the second point.
SOUR:POW:CORR:TCO:INS2 1.5 GHz, -7.5, DUT
Insert a third point as point no. 2 in the list.
SOUR:POW:CORR:TCO:COUN? DUT
Query the number of points. The response is 3.
SOUR:POW:CORR:TCO ON, DUT
Enable the use of two-port transmission coefficients.
SOUR:POW:CORR:TCO:FEED 'Trc1', DUT
Replace the previous 3 points by the trace points of the default
trace "Trc1".
MMEM:STOR:CORR:TCO 'C:
\Users\Public\Documents\Rohde-Schwarz\Vna
\PowerMeterCorr\Test.pmcl', DUT
Store the power loss list to the power meter correction list file
Test.pmcl in the directory
C:\Users\Public\Documents\Rohde-Schwarz\Vna
\PowerMeterCorr.
MMEM:LOAD:CORR:TCO 'C:\Users\Public\Documents
\Rohde-Schwarz\Vna\PowerMeterCorr\Test.pmcl',
DUT
Re-load the stored power meter correction list file.
SOUR:POW:CORR:TCO:DEL1, DUT
Delete the first point in the list.
SOUR:POW:CORR:TCO:DEL:ALL, DUT; :SOUR:POW:CORR:
TCO? DUT
Clear the entire list. Query whether the transmission coefficients
are still taken into account. The response is 0.
MMEM:LOAD:CORR:TCO 'C:\Users\Public\Documents
\Rohde-Schwarz\Vna\Traces\Test.s2p', 'S21', DUT
Load a power loss list from a previously created 2-port Touch-
stone file.
Manual operation:
See "Two Port Config..." on page 530
7.3.15.5 SOURce:POWer:CORRection... (other)
The SOURce:POWer:CORRection... commands control a scalar source power cali-
bration.
SOURce<Ch>:POWer:CORRection[:ACQuire]................................................................1359
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]:VERification:RESult?.................... 1360
SOURce<Ch>:POWer:CORRection:DATA...................................................................... 1361
SOURce<Ch>:POWer:CORRection:DATA:PORT<PhyPt>................................................1361
SOURce<Ch>:POWer:CORRection:DATA:PARameter<Wv>?.......................................... 1362
SOURce<Ch>:POWer<PhyPt>:CORRection:DATA:PARameter<Wv>:COUNt?...................1363
SOURce<Ch>:POWer:CORRection:GENerator<Gen>[:STATe].........................................1364
SOURce<Ch>:POWer<PhyPt>:CORRection:IMODulation:LTONe[:ACQuire]......................1364
SCPI command reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals