EasyManua.ls Logo

R&S ZNB Series - 7.3.1.4 CALCulate:EYE...

R&S ZNB Series
1611 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNB/ZNBT
858User Manual 1173.9163.02 ─ 62
Note: The skew measurement can only be performed if the following conditions are
met for both the active trace and the reference trace:
the trace format is real (CALCulate<Chn>:FORMat REAL)
Time Domain is enabled (CALCulate<Chn>:TRANsform:TIME:STATe ON)
the Low Pass Step time domain transform is used CALCulate<Chn>:
TRANsform:TIME[:TYPE] LPASs and CALCulate<Chn>:TRANsform:TIME:
STIMulus STEP)
The latter, in turn, requires the stimulus grid to be harmonic. This can be achieved, for
example, using [SENSe<Ch>:]HARMonic:AUTO ON.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> *RST: OFF
Options: R&S ZNB/ZNBT-K20
Manual operation: See "Skew" on page 709
CALCulate<Chn>:DTIMe:TARGet <SkewRefTrace>
Selects the reference trace for the skew measurement.
See CALCulate<Chn>:DTIMe:STATe for conditions on both the active and the refer-
ence trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<SkewRefTrace> Reference trace, identified by its name (within channel <Chn>).
Options: R&S ZNB/ZNBT-K20
Manual operation: See "Reference Trace" on page 709
7.3.1.4 CALCulate:EYE...
Defines the properties and retrieves the results of the eye diagram measurement provi-
ded with the Extended time domain analysis option R&S ZNB/ZNBT-K20.
CALCulate<Chn>:EYE:DUT:MODE..................................................................................860
CALCulate<Chn>:EYE:EMPHasis:CURSor:POST<1|2>.....................................................860
CALCulate<Chn>:EYE:EMPHasis:CURSor:PRE...............................................................860
CALCulate<Chn>:EYE:EMPHasis:STATe......................................................................... 861
CALCulate<Chn>:EYE:EQUalization:CTLE:DC.................................................................861
CALCulate<Chn>:EYE:EQUalization:CTLE:POLE<1|2>.....................................................861
CALCulate<Chn>:EYE:EQUalization:CTLE:ZERO.............................................................862
CALCulate<Chn>:EYE:EQUalization:STATe......................................................................862
CALCulate<Chn>:EYE:INPut:BPATtern:TYPE...................................................................863
CALCulate<Chn>:EYE:INPut:DRATe................................................................................863
SCPI command reference

Table of Contents

Related product manuals