EasyManua.ls Logo

R&S ZNB Series - 5.16.2.9 Rise Time tab

R&S ZNB Series
1611 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI reference
R&S
®
ZNB/ZNBT
707User Manual 1173.9163.02 ─ 62
Save / Load Mask Configuration
Opens a dialog that allows to save/load the mask test configuration to/from a 7bit
ASCII file (*.mask). An example is shown below:
[Mask Top]
Active = true
Width Top = 9e-009
Offset Top = 6
[Mask Bottom]
Active = true
Width Bottom = 9e-009
Offset Bottom = 6
[Mask Polygon]
Active = true
Shape = Octagon
Main Width = 9e-009
Minor Width = 8e-009
Main Height = 9
Minor Height = 8
[MASK CENTER]
Horizontal = 0
Vertical = 0
[TEST SETTING]
Fail Condition = Samples (Percent)
Violation Tolerance = 1
Violation Percent = 10
Remote command:
MMEMory:LOAD:EYE:MASK
MMEMory:STORe:EYE:MASK
Automatic Mask Generation
The "Automatic Mask Generation" function attempts to size the polygon, top, and bot-
tom masks in a reasonable manner based on the current eye measurement settings.
Remote command:
CALCulate<Chn>:EYE:MASK AUTO
5.16.2.9 Rise Time tab
The "Rise Time" tab allows you to enable and configure the Rise Time Measurement.
Access: System – [Applic] > "TDR" > "Rise Time"