Parameters:
<ThresholdEnum|LowerThreshold>T1_9: 10–90% rise time
T2_8: 20–80% rise time
<integer>: Lower rise time threshold as integer percentage
*RST: T1_9:
<UpperThreshold> Upper rise time threshold as integer percentage
Options: R&S ZNB/ZNBT-K20
Manual operation: See "Start Value / Stop Value" on page 708
7.3.1.15 CALCulate... (other)
CALCulate:CLIMits:FAIL?............................................................................................. 1059
CALCulate<Chn>:DLINe...............................................................................................1060
CALCulate<Chn>:DLINe:STATe.....................................................................................1060
CALCulate<Chn>:FORMat............................................................................................1060
CALCulate<Chn>:FORMat:WQUType............................................................................1062
CALCulate<Chn>:GDAPerture:SCOunt..........................................................................1062
CALCulate<Chn>:IAVerage:MODE................................................................................ 1062
CALCulate<Chn>:IAVerage[:STATe]............................................................................... 1063
CALCulate<Chn>:PHOLd............................................................................................. 1063
CALCulate:PIPelining:DATA?........................................................................................ 1064
CALCulate<Chn>:SMOothing:APERture........................................................................ 1064
CALCulate<Chn>:SMOothing[:STATe]............................................................................1065
CALCulate:TDVSwr[:STATe]..........................................................................................1065
CALCulate:CLIMits:FAIL?
Returns a 0 or 1 to indicate whether or not a global, composite limit check on several
traces has failed.
Since V2.20 of the R&S ZNB/ZNBT FW the result is automatically recalculated when-
ever a relevant setting is changed, i.e. a subsequent query will return the updated limit
violation state.
Example:
*RST; CALC:LIM:CONT 1 GHZ, 2 GHZ
Define an upper limit line segment in the stimulus range
between 1 GHz and 2 GHz, using default response values.
CALC:LIM:STAT ON; FAIL?
Switch the limit check on and query the result.
CALC:CLIM:FAIL?
Query the result for the composite limit check. As only one trace
is tested, the response should be equal to the previous
response.
Usage: Query only
Manual operation: See "Global Check" on page 376