GUI reference
R&S
®
ZNB/ZNBT
516User Manual 1173.9163.02 ─ 62
[SENSe:]CORRection:CKIT:DMODe
[SENSe:]CORRection:CKIT:DELete
Query connector types and calibration kits:
[SENSe<Ch>:]CORRection:CONNection:CATalog?
[SENSe:]CORRection:CKIT:CATalog?
Import Cal Kit... / Export Cal Kit...
The buttons below the "Connector Type" list are used to store cal kit data to a file and
to reload previously stored cal kit files. By default, calibration kit files are stored in the
C:\Users\Public\Documents\Rohde-Schwarz\Vna\Calibration directory;
see Chapter 4.5.2.4, "Cal kit Files", on page 179.
Remote command:
MMEMory:LOAD:CKIT
MMEMory:STORe:CKIT
Kit Standards dialog
The "Kit Standards" dialog shows the calibration standards in a selected calibration kit.
It is also used to modify the contents of a user-defined kit.
Related information
Refer to the following sections:
●
Chapter 4.5.2.4, "Cal kit Files", on page 179
●
Chapter 5.11.2.2, "Calibration Kits dialog", on page 514
●
Chapter 4.5.2.1, "Calibration standard types", on page 175
Access: Calibration Kits dialog > "Standards..."
Cal softtool