GUI reference
R&S
®
ZNB/ZNBT
790User Manual 1173.9163.02 ─ 62
●
The ICC does not support hot plugging of ICUs, so connect the ICUs to the ICC
before connecting the ICC to the R&S ZNB/ZNBT.
●
For background information, see Chapter 4.5.5.4, "Inline calibration", on page 186.
Configured Devices
Each ICC R&S ZN-Z30 is listed with its connected Inline Calibration Units (ICUs)
R&S ZN-Z32 and R&S ZN-Z33.
Remote command:
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:SUBModule:ALL?
5.18.5 External Ports tab
Allows you to set up switch matrices and to redefine physical ports.
Background information
Refer to:
●
Chapter 4.7.29, "External switch matrices", on page 269
●
Chapter 4.3.1.2, "Redefined S-parameters", on page 127
5.18.5.1 Controls on the External Ports tab
The buttons in the "External Ports" panel open the following dialogs:
●
"Define Ports...", see Chapter 5.18.5.2, "Define Physical Ports dialog",
on page 791
●
"Switch Matrix...", see Chapter 5.18.5.3, "External Matrices dialog", on page 791
●
"RF Connections...", see Chapter 5.18.5.5, "Switch Matrix RF Connections dialog",
on page 796
"RF Connections" is enabled if at least one switch matrix is configured.
Optimization
Allows you to select between different switch matrix routing optimizations:
●
"Speed" – switch as little as possible
●
"Precision" – always use the best possible route in terms of quality ("priority")
Setup softtool