EasyManua.ls Logo

R&S ZNB Series - 5.7.4 Multiple Peak tab; 5.7.4.1 Controls on the Multiple Peak tab

R&S ZNB Series
1611 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI reference
R&S
®
ZNB/ZNBT
402User Manual 1173.9163.02 ─ 62
Remote command:
CALCulate<Chn>:MARKer<Mk>:EXCursion
CALCulate<Chn>:MARKer<Mk>:REFerence:EXCursion
CALCulate<Chn>:MARKer<Mk>:EXCursion:STATe
CALCulate<Chn>:MARKer<Mk>:REFerence:EXCursion:STATe
5.7.4 Multiple Peak tab
"Multiple Peak" search allows you to find multiple local minima/maxima at once.
Background information
Refer to "Basic marker search functions" on page 110.
5.7.4.1 Controls on the Multiple Peak tab
Max / Min
Sets up to 10 markers to the highest maxima or lowest minima in the configured Eval
Range. If a complex trace format is active (e.g. a polar diagram), the markers are set to
the measurement points with the maximum or minimum magnitude.
The required markers are created/deleted as needed.
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute MMAXimum | MMINimum
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:EXECute MMAXimum |
MMINimum
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult?
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:RESult?
Eval Range
Opens the Evaluation Range dialog that allows you to set the domain for the multiple
peak search. A modified domain takes effect the next time Max / Min is used.
Marker softtool

Table of Contents

Related product manuals