<PhyPt> Test port number of the analyzer. This suffix is ignored because
the selected frequency applies to all source ports in the active
channel.
Parameters:
<FixedFreq> Fixed stimulus and analyzer frequency.
*RST: 1 GHz
Default unit: Hz
Example:
FUNC "XTIMe:POW:A1"
Activate a time sweep and select the wave quantity a
1
as mea-
sured parameter for channel and trace no. 1.
FREQ:CW 100MHz
Set the CW frequency to 100 MHz.
Manual operation: See "CW Frequency" on page 415
7.3.15.2 SOURce:GROup...
The SOURce:GROup... commands define the group of (un-)used logical ports.
SOURce<Ch>:GROup<Grp>........................................................................................ 1337
SOURce<Ch>:GROup<Grp>:CLEar.............................................................................. 1338
SOURce<Ch>:GROup:COUNt?.....................................................................................1338
SOURce<Ch>:GROup<Grp>:DPORt:COUNt..................................................................1339
SOURce<Ch>:GROup<Grp>:NAME.............................................................................. 1339
SOURce<Ch>:GROup<Grp>:PORDer........................................................................... 1339
SOURce<Ch>:GROup<Grp>:PORTs............................................................................. 1340
SOURce<Ch>:GROup<Grp>:PPORt<PhyPort>:DPORt................................................... 1340
SOURce<Ch>:GROup<Grp>:PPORts............................................................................1341
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:CONDition......................................1342
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:MOFFset:BWFactor........................ 1344
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:MOFFset:DVALue........................... 1344
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:MOFFset:MODE............................. 1345
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:STATe............................................ 1345
SOURce<Ch>:GROup<Grp> <arg0>, <arg1>
In "set direction" this command defines the set of active logical ports (the "master
group") as the continuous range from <log_port1> to <log_port2>, i.e. logical
ports outside this range are disabled.
In "get direction" it returns the minimum and maximum enabled logical port for the rela-
ted group.
See SOURce<Ch>:LPORt<LogPt> on how to define logical ports.
Note:
Setting SOURce<Ch>:GROup or SOURce<Ch>:GROup<Grp>:PORTs automatically
dissolves the port groups created using SOURce<Ch>:GROup<Grp>:PPORts.