GUI reference
R&S
®
ZNB/ZNBT
759User Manual 1173.9163.02 ─ 62
The decimal values have the following effect:
●
0 means that no output signals are enabled at any of the pins 8, 9, 10, 11 and 16,
17, 18, 19.
●
1 enables the output signal at pin 8. The signal is switched on while a measure-
ment sweep is running in the selected channel. All other signals are inactive.
●
2 enables the output signal at pin 9.
●
3 enables the output signals at pins 8 and 9.
●
...
●
255 enables the output signals at all pins. See also "Pin 16 - 19" on page 759.
Remote command:
OUTPut<Ch>:UPORt[:VALue]
Pin 16 - 19
Selects the control mechanism for the signals at pins 16 to 19 of the User Port connec-
tor.
●
"Channel Bits": Signals are controlled by channel bits 4 to 7. No drive port indica-
tion at the User Port connector.
●
"Drive Ports": Signals indicate the active drive ports. The number of active chan-
nel bits is reduced to 4 (pins 8 to 11).
Remote command:
OUTPut:UPORt:ECBits
Advanced tab
Collects several advanced settings.
Geometric Calculation of Bandfilter Center
Defines how bandfilter searches calculate the center frequency of the passband or
stopband (see "Bandfilter search" on page 110).
If "Geometric Calculation of Bandfilter Center" is checked, the geometric mean of the
lower band edge and upper band edge frequencies is used, otherwise their arithmetic
mean.
Setup softtool