Command reference
R&S
®
ZNB/ZNBT
1368User Manual 1173.9163.02 ─ 62
SOURce:POWer:CORRection:PSELect <CalPower>
Defines how to define the source power the R&S ZNB/ZNBT uses to perform the first
calibration sweep of the source power calibration ("Reference Receiver Cal Power").
Parameters:
<CalPower> CPOWer | PPOWer
CPOWer
Calculated port power result
PPOWer
Predefined port power value, see SOURce:POWer:
CORRection:PPOWer
*RST: CPOWer
Example: See SOURce<Ch>:POWer:CORRection[:ACQuire].
Manual operation: See "Reference Receiver Cal Power" on page 528
SOURce<Ch>:POWer<PhyPt>:CORRection:STATe <Boolean>
Enables or disables the source power calibration for channel <Ch> and for port number
<PhyPt>. The setting command is disabled unless a source power calibration for the
analyzer port has been performed (SOURce<Ch>:POWer:CORRection[:ACQuire]).
The query always returns a result.
To enable or disable a source power calibration for an external generator use
SOURce<Ch>:POWer:CORRection:GENerator<Gen>[:STATe].
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt> Calibrated port number
Parameters:
<Boolean> Enables (ON) or disables (OFF) the source power calibration for
port number <PhyPt>.
*RST: OFF
Example: See SOURce<Ch>:POWer:CORRection[:ACQuire].
Manual operation: See "Port Overview" on page 534
7.3.15.6 SOURce:POWer... (other)
The SOURce:POWer... commands define the power of the internal signal source.
SOURce<Ch>:POWer<PhyPt>:CORRection:GENerator<Gen>:LEVel:OFFSet...................1369
SOURce<Ch>:POWer<PhyPt>:CORRection:LEVel:OFFSet............................................. 1370
SOURce<Ch>:POWer<PhyPt>:GENerator<Gen>:OFFSet............................................... 1370
SOURce<Ch>:POWer<PhyPt>:GENerator<Gen>:PERManent[:STATe]............................. 1371
SOURce:POWer:GENerator:SDELay.............................................................................1372
SOURce<Ch>:POWer<PhyPt>:GENerator<Gen>:STATe................................................. 1372
SCPI command reference