Manual operation: See "Apply" on page 471
[SENSe<Ch>:]CORRection:COLLect:SCONnection<PhyPt> <Type>[, <Gender>]
Selects a connector type at a specified port <PhyPt> and its gender. In contrast to
[SENSe<Ch>:]CORRection:COLLect:CONNection<PhyPt>, this command uses
a string variable to identify the connector type.
Suffix:
<Ch>
.
Channel number of the calibrated channel
<PhyPt> Port numbers of the analyzer. If the analyzer is set to use the
same connectors at all ports ([SENSe<Ch>:]CORRection:
COLLect:CONNection:PORTs ALL), then a change of a con-
nector type is valid for all ports. The gender of the connectors
can still be different.
Parameters:
<Type> Connector type (string parameter) of the connectors. See "Cal
kit naming conventions" under [SENSe:]CORRection:CKIT:
<ConnType>:SELect.
*RST: 'N 50 Ohm',FEM for all ports.
<Gender> MALE | FEMale
Gender of the connectors. The gender designation is not neces-
sary (and ignored) for sexless connector types.
Example:
*RST; :CORR:COLL:SCON1 'N 75 Ohm', MALE; SCON4?
Change the connector type at port 1 from 'N 50 Ohm', FEM to 'N
75 Ohm', MALE. The connector type at the other ports is also
changed to N 75 Ohm, however, the gender (female) is main-
tained. CORR:COLL:SCON4? returns 'N 75 Ohm', FEM.
Manual operation: See "Connector / Gender" on page 467
7.3.14.7 [SENSe:]CORRection:METas
Using [SENSe:]CORRection:METas commands you can define additional properties
of a METAS Calibration.
To perform a METAS calibration, use the standard commands for manual calibration
(see Chapter 7.3.14.8, "[SENSe:]CORRection... (other)", on page 1245). For the cali-
bration setup, use [SENSe<Ch>:]CORRection:COLLect:METHod:DEFine with
METAS calibration types MTOSm | MUOSm | MOSM | MTRL.
SCPI command reference