Command reference
R&S
®
ZNB/ZNBT
1241User Manual 1173.9163.02 ─ 62
Example:
CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard
as calibration type.
CORR:COLL:SAVE:SEL:DEF
Calculate a dummy system error correction for the normalization
at port 1. The dummy system error correction provides the
reflection tracking error term 'REFLTRACK'.
CORR:CDAT? 'REFLTRACK',1,0
Query the dummy system error correction term. The response is
a 1 (written as 1,0 for the real and imaginary part) for each
sweep point (no attenuation and no phase shift between the
analyzer and the calibration plane).
INIT:CONT OFF; :INIT; *WAI
Stop the sweep to ensure correct transfer of calibration data.
CORR:CDAT 'REFLTRACK',1,0,<ASCII_data>
Replace the dummy system error correction term with your own
correction data, transferred in ASCII format.
INIT:CONT ON
Restart the sweep in continuous mode.
Usage:
Event
Manual operation: See "Apply" on page 471
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected[:DUMMy]
Calculates the system error correction data from the acquired measurement results
([SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected), stores it and
applies it to the calibrated channel(s). To avoid incompatibilities, older system error cor-
rection data is deleted unless it has been transferred into a cal pool (MMEMory:
STORe:CORRection <Ch>, '<file_name>').
Suffix:
<Ch>
.
One of the calibrated channels
If [SENSe:]CORRection:COLLect:CHANnels:ALL is ON
this suffix can be omitted
Example:
CORRection:COLLect:METHod:DEFine 'Test',RSHort,
1
Select a one-port normalization at port 1 with a short standard
as calibration type.
CORRection:COLLect:CHANnels:ALL ON
Enable calibration in all channels of the active recall set.
CORRection:COLLect:SELected SHOR,1
Measure a short standard connected to port 1 and store the
measurement results of this standard.
CORRection:COLLect:SAVE:SELected
Calculate the system error correction data and apply it to all
channels of the active recall set.
Usage:
Event