Command reference
R&S
®
ZNB/ZNBT
1227User Manual 1173.9163.02 ─ 62
[SENSe:]CORRection:COLLect:CHANnels:ALL <Boolean>
Enables calibration of all channels in the active recall set. The command is valid for the
following calibration methods:
Manual system error correction
Automatic system error correction (cal unit)
Manual SMARterCal
Automatic SMARTerCal (cal unit)
A scalar power calibration is not affected.
Parameters:
<Boolean> ON calibrate all channels.
OFF calibrate the active channel only.
Example: See [SENSe<Ch>:]CORRection:COLLect:SAVE:
SELected[:DUMMy]
Manual operation: See "Calibrate all Channels" on page 465
[SENSe:]CORRection:COLLect:CHANnels:MCTYpes <Boolean>
Toggles the Multiple Calibration TYpes mode that allows you to calibrate a subset of
the available channels using channel-specific ports and calibration types (see Chap-
ter 4.5.8, "Parallel calibration of multiple channels", on page 201).
Enable the MCTYpes mode before defining the calibrations to be performed using
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine. Disable the MCTYpes
mode after the calibrations were saved using [SENSe<Ch>:]CORRection:
COLLect:SAVE:SELected[:DUMMy]. The latter must be called only once, with <Ch>
referring to one of the channels to be calibrated (see the example below).
Parameters:
<Boolean>
SCPI command reference