EasyManuals Logo

R&S ZNB Series User Manual

R&S ZNB Series
1611 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #419 background imageLoading...
Page #419 background image
GUI reference
R&S
®
ZNB/ZNBT
419User Manual 1173.9163.02 ─ 62
Remote command:
CALCulate<Chn>:TRANsform:TIME:STARt
CALCulate<Chn>:TRANsform:TIME:STOP
CALCulate<Chn>:TRANsform:TIME:CENTer
CALCulate<Chn>:TRANsform:TIME:SPAN
Distance Start / Distance Stop / Distance Center / Distance Span
Defines the display range for the time domain trace in distance representation (see "
Time / Distance" on page 419).
"Distance Start" and "Distance Stop" or "Distance Center" and "Distance Span" are
alternative settings.
Remote command:
n.a.
See Time Start / Time Stop / Time Center / Time Span.
Time / Distance
"Time" and "Distance" switch over between the x-axis scaling in time units or distance
units.
The interpretation of time and distance depends on the measurement type. For reflec-
tion measurements (S-parameters S
ii
or ratios with equal port indices), the time axis
represents the propagation time of a signal from the source to the DUT and back. For
transmission measurement, it represents the propagation time from the source through
the device to the receiver.
The distance between the source and the DUT is calculated from the propagation time,
the velocity of light in the vacuum, and the velocity factor of the receiving port:
Distance = <Time> * c
0
* <Velocity Factor> for transmission measurements
Distance = 1/2 * <Time> * c
0
* <Velocity Factor> for reflection measurements. The
factor 1/2 accounts for the return trip from the DUT to the receiver.
The velocity factor of the receiving port can be defined using Channel – [Offset Embed]
> "Offset" (see Chapter 5.13.2.1, "Controls on the Offset tab", on page 616).
Remote command:
CALCulate<Chn>:TRANsform:TIME:XAXis
5.9 Power Bw Avg softtool
The "Power Bw Avg" softtool allows you to configure the signal power, to set up the IF
signal processing, and to configure the averaging logic.
Access: Channel – [Pwr Bw Avg] hardkey
5.9.1 Power tab
The "Power" tab allows you to configure the signal power. It is identical to the "Stimu-
lus" > "Power" tab; see Chapter 5.8.2, "Power tab", on page 415.
Power Bw Avg softtool

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals