GUI reference
R&S
®
ZNB/ZNBT
312User Manual 1173.9163.02 ─ 62
DUT
Balanced
port
Balanced
port
Log.
NWA
port
Physical port c
Logical ports j,k
(single-ended)
Physical port d
Logical port i
(balanced)
Physical port a
Physical port b
Imbalance 1 / Imbalance 2
Imbalance 1 selects the balanced logical port i as the input, the single-ended logical
ports j and k as the output and calculates the following imbalance parameter:
Imb
jk-i
= –(S
ca
– S
da
)/(S
cb
– S
db
)
Imbalance 2 selects the balanced logical port i as the output, the single-ended logical
ports j and k as the input and calculates the following imbalance parameter:
Imb
i-jk
= –(S
ac
– S
ad
)/(S
bc
– S
bd
)
Remote command:
CALCulate<Ch>:PARameter:MEASure "<Trace_Name>", "IMB<i>-<j><k>"
| "IMB<j><k>-<i>"
CALCulate<Ch>:PARameter:SDEFine "<Trace_Name>", "IMB<i>-<j><k>"
| "IMB<j><k>-<i>"
5.2.9 Stability tab
Selects one of the three two port stability factors K, μ
1
or μ
2
as measured quantities. A
typical application of stability factors is to assess the stability of an amplifier. Stability
factors cannot be calculated in balanced port configurations.
Background information
Refer to Chapter 4.3.7, "Stability factors", on page 142.
Meas softtool

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals