EasyManua.ls Logo

R&S ZNB Series - Step 3: Calibration; 5.11.1.12 Cal Validation dock widget; Basic mode

R&S ZNB Series
1611 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI reference
R&S
®
ZNB/ZNBT
509User Manual 1173.9163.02 ─ 62
Start
Start Step 3: Calibration.
Step 3: Calibration
Allows you to acquire error correction data for every required port (pair) and calibration
standard, plus power correction data from the power meter.
Same functionality as for the single calibration version of the wizard (see "Step 3: Cali-
bration" on page 468).
The order of the system error correction and power calibration sweeps is arbitrary,
but make sure that you always connect the proper equipment.
The possibility to "share" measurements between the calibrations and the "Reduced
Through" logic help to keep the number of measurements as small as possible (see
Chapter 4.5.1.11, "Full n-Port calibration with reduced number of Through connec-
tions", on page 173).
5.11.1.12 Cal Validation dock widget
The controls on the "Cal Validation" dock widget allow you to set up and run a Calibra-
tion validation in basic or advanced mode.
Basic mode
In basic mode, the validation runs automatically, with autodetected ports and the set-
tings specified in Advanced mode (default settings after a system [Preset]).
Use "Detect Port Assignment" to detect the current port assignment (after reconnecting
the calibration unit).
Manual port assignment is not supported.
Validate
Runs the validation for the current port assignment. The validation proceeds through
the connected ports p {1, ..., 4} in ascending order, measuring S
pp
for the Open,
Short and Match standard.
Cal softtool

Table of Contents

Related product manuals