Remote control
R&S
®
ZNB/ZNBT
832User Manual 1173.9163.02 ─ 62
Bit No. Meaning
4 Execution error
This bit is set if a received command is syntactically correct, but cannot be performed for
other reasons. An error message with a number between 200 and 300, which describes
the error in greater detail, is entered into the error queue.
5 Command error
This bit is set if a command which is undefined or syntactically incorrect is received. An
error message with a number between -100 and -200, which describes the error in greater
detail, is entered into the error queue.
6 User request
This bit is set when the instrument is switched over to manual control or when a user-
defined softkey is used (see SYSTem:USER:KEY).
7 Power On (supply voltage on)
This bit is set when the instrument is switched on.
Related common commands
The Event Status Register (ESR) can be queried using ESR?. The Event Status Enable
(ESE) register can be set using the command *ESE and read using *ESE?.
See also Common commands.
6.5.3.4 STATus:OPERation
The STATus:OPERation register contains conditions which are part of the instrument's
normal operation. The analyzer does not use the STATus:OPERation register.
6.5.3.5 STATus:QUEStionable
The STATus:QUEStionable register indicates whether the acquired data is of ques-
tionable quality and monitors hardware failures of the analyzer. It can be queried using
the commands STATus:QUEStionable:CONDition? or
STATus:QUEStionable:EVENt?.
Bit No. Meaning
9 INTegrity register summary
This bit is set if a bit is set in the STATus:QUEStionable:INTegrity register and the associ-
ated ENABle bit is set to 1.
10 LIMit register summary
This bit is set if a bit is set in the STATus:QUEStionable:LIMit1 register and the associated
ENABle bit is set to 1.
STATus:QUEStionable:LIMit<1|2>
The STATus:QUEStionable:LIMit<1|2> registers indicate the result of the limit
check. They can be queried using the commands
STATus:QUEStionable:LIMit<1|2>:CONDition? or
Status reporting system

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals