Command reference
R&S
®
ZNB/ZNBT
1193User Manual 1173.9163.02 ─ 62
7.3.14.4 [SENSe:]CORRection:CKIT...
The [SENSe:]CORRection:CKIT... commands deal with calibration kits and cal kit
data. The calibration kits are distinguished by their names (<CalkitName>), the optional
labels (<label>) can be used to carry information about the calibration standard.
In order to handle several identical calibration kits with different serial numbers use the
commands of Chapter 7.3.14.5, "[SENSe:]CORRection:CKIT... with Labels",
on page 1201.
[SENSe:]CORRection:CKIT:<StandardType>.................................................................. 1193
[SENSe:]CORRection:CKIT:CATalog?............................................................................ 1197
[SENSe:]CORRection:CKIT:DELete............................................................................... 1197
[SENSe:]CORRection:CKIT:DMODe.............................................................................. 1197
[SENSe:]CORRection:CKIT:LABel..................................................................................1198
[SENSe:]CORRection:CKIT:SELect................................................................................1198
[SENSe:]CORRection:CKIT:<ConnType>:SELect............................................................ 1199
[SENSe:]CORRection:CKIT:STANdard:CATalog?............................................................ 1200
[SENSe:]CORRection:CKIT:STANdard:DATA?................................................................ 1200
[SENSe:]CORRection:CKIT:<StandardType> <ConnType>, <CalKitName>,
<StandardLabel>, <MinFreq>, <MaxFreq>[, <DelayParam>, <Loss>, <Z0>[,
<C0>, <C1>, <C2>, <C3>, <L0>, <L1>, <L2>, <L3>[, OPEN | SHORt | MATCh |
<Resistance>[, <Port1>[, <Port2>]]]]]
[SENSe:]CORRection:CKIT:<StandardType>? <ConnType>, <CalKitName>[,
<Port1>[, <Port2>]]
Defines the parameters of a (possibly non-ideal) 1 port or 2-port calibration standard
<StandardType> within a particular cal kit. Depending on the standard type, only a
subset of the parameters is used; see Table 7-13
Note
As the query does not reflect the delay mode specified using [SENSe:]CORRection:
CKIT:DMODe, it is deprecated and only available for backward compatibility reasons.
Use [SENSe:]CORRection:CKIT:STANdard:DATA?? instead.
Suffix:
<StandardType>
.
<string>
Standard type
For one-port standards, the first character denotes the gender,
for transmission standards the first two characters denote the
genders on both ends, e.g. MOPen for a male Open standard or
FFTHrough for a Through standard with female connectors.
For a complete list of standards, refer to Table 7-15.
Parameters:
<StandardLabel> Additional string label for the standard, typically the standard's
serial number
SCPI command reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals