Command reference
R&S
®
ZNB/ZNBT
1199User Manual 1173.9163.02 ─ 62
Example:
MMEM:LOAD:CKIT 'C:\Users\Public\
Documents\Rohde-Schwarz\Vna\
Calibration\Kits\New Kit 1.calkit'
Load the previously created cal kit file New Kit 1.calkit
from the default cal kit directory.
CORR:CKIT:SEL 'N 50 Ohm', 'New Kit 1'
Assign the imported kit to the N 50 Ω connector type, assuming
that the cal kit name stored in New Kit 1.calkit reads New
Kit 1.
Manual operation: See "Cal Kit" on page 467
[SENSe:]CORRection:CKIT:<ConnType>:SELect
<CalKitName>
Selects the calibration kit to be used for a specified connector type <ConnType>. The
kit is identified by its name.
Tip: For connector types with arbitrary, user-defined names you can use the command
[SENSe:]CORRection:CKIT:SELect.
Cal kit naming conventions
Calibration kit names must be entered as string parameters. The string contains the cal
kit name used in the calibration dialogs (e.g. "Calibration Presettings"); a "Ω" in the
name must be replaced by Ohm, e.g.:
NewKit1 denotes the user-defined calibration kit "NewKit1".
N 50 Ohm Ideal Kit denotes the "N 50 Ω Ideal Kit".
ZV-Z21 typical denotes the cal kit "ZV-Z21 typical".
Suffix:
<ConnType>
.
Connector type, one of the following identifiers:
N50, N75 N 50 Ω or N 75 Ω connectors
PC7, PC35, PC292 PC 7, PC 3.5 or 2.92 mm connectors
USER<no> user-defined connectors UserConn1, UserConn2
SMA user-defined connector type SMA ...
Parameters:
<CalKitName> String parameter containing the name of a calibration kit availa-
ble on the analyzer. See "Cal kit naming conventions " above.
*RST: n/a (A *RST does not change the assignment
between connector types and calibration kits.)
Example:
MMEM:LOAD:CKIT 'C:\Users\Public\
Documents\Rohde-Schwarz\Vna\
Calibration\Kits\New Kit 1.calkit'
Load the previously created cal kit file New Kit 1.calkit
from the default cal kit directory.
CORR:CKIT:N50:SEL 'New Kit 1'
Assign the imported kit to the N 50 Ω connector type, assuming
that the cal kit name stored in New Kit 1.calkit reads New
Kit 1.
SCPI command reference