Concepts and features
R&S
®
ZNB/ZNBT
250User Manual 1173.9163.02 ─ 62
4.7.8 Eazy de-embedding based on IEEE 370
Option R&S ZNB-K210/R&S ZNBT-K210
EaZy deembedding (EZD) is an implementation of the IEEE algorithm for generating
deembedding files from a symmetrical 2x thru measurement.
Option R&S ZNB/ZNBT-K210 allows you to model a test fixture using the EZD fixture
modeling tool and to deembed selected ports using the generated touchstone files.
The EZD fixture modeling proceeds in the following steps:
1. Measure one or more test coupons for the related fixture.
2. Run the EZD tool to calculate the Touchstone files modeling the test fixture.
The calculated sNp files can then be used to deembed the DUT at selected ports.
Figure 4-24: Fixture modeling: EZD – Single Ended Ports dialog