Loading...
Command reference
R&S
®
ZNB/ZNBT
1186User Manual 1173.9163.02 ─ 62
PROGram[:SELected]:RETVal?
Queries the return value of an application or process started via PROGram[:
SELected]:EXECute.
This will only be successful if preceded by a PROGram[:SELected]:WAIT? query
(see PROGram[:SELected]:WAIT on page 1186).
Example: See PROGram[:SELected]:EXECute
Usage: Query only
PROGram[:SELected]:WAIT
Locks command execution from the current controller program while a program started
via PROGram[:SELected]:EXECute is running. The analyzer does not execute any
further commands or queries until the program is stopped or paused.
Use PROGram[:SELected]:WAIT? before trying to retrieve the return value of the
executed program (PROGram[:SELected]:RETVal?).
Example:
See PROGram[:SELected]:EXECute
7.3.14 [SENSe:] commands
The [SENSe:]... commands affect the receiver settings of the R&S ZNB/ZNBT.
7.3.14.1 [SENSe:]AVERage...
The [SENSe:]AVERage... commands set sweep averaging parameters. The sweep
average is a noise-reduction technique which consists of calculating each measure-
ment point as an average of the same measurement point over several consecutive
sweeps.
In contrast to the sweep count (for single sweep mode, [SENSe<Ch>:]SWEep:COUNt
), averaging is always channel-specific. Both features are completely independent from
each other.
[SENSe<Ch>:]AVERage[:STATe]................................................................................... 1186
[SENSe<Ch>:]AVERage:CLEar..................................................................................... 1187
[SENSe<Ch>:]AVERage:COUNt.................................................................................... 1187
[SENSe<Ch>:]AVERage:MODE.....................................................................................1187
[SENSe<Ch>:]AVERage[:STATe] <Boolean>
Enable or disable the sweep average.
Suffix:
<Ch>
.
Channel number