Command reference
R&S
®
ZNB/ZNBT
1321User Manual 1173.9163.02 ─ 62
Example:
*RST; SOUR:POW 5
Define a base channel power of +5 dBm. This power is also
used as an (initial) power for all test ports and external genera-
tors.
SOUR:POW1:OFFS -5, CPAD
Change the source power at test port no. 1 to 0 dB without
affecting the power at the remaining ports.
SWE:AXIS:POW 'Port 1; Source'
Select the source signal at port 1 as the reference signal for
power definitions.
SOUR:POW -5
Shift all powers by -5 GHz.
SWE:AXIS:POW 'Port 2; Source'
Select the source signal at port 2 as the reference signal for
power definitions.
SOUR:POW?
Query the power at test port 2. The response is 0.
Manual operation:
See "Stimulus Axis – Frequency / Power" on page 561
[SENSe<Ch>:]SWEep:COUNt <Sweep>
Defines the channel-specific number of sweeps to be measured and buffered in single
sweep mode (INITiate<Ch>:CONTinuous OFF). Use [SENSe:]SWEep:COUNt:
ALL to define the sweep count for all channels.
Suffix:
<Ch>
.
Channel number
Parameters:
<Sweep> Number of consecutive sweeps to be measured and buffered.
Range: 1 to 100000
*RST: 1
Example: See CALCulate<Chn>:DATA:NSWeep:FIRSt?
Manual operation: See "Sweeps" on page 442
[SENSe:]SWEep:COUNt:ALL <Sweep>
Defines the number of sweeps to be measured in single sweep mode
(INITiate<Ch>:CONTinuous OFF). The setting is applied to all channels. Use
[SENSe<Ch>:]SWEep:COUNt to define the sweep count for a single channel.
Setting parameters:
<Sweep> Number of consecutive sweeps to be measured.
Range: 1 to 100000
*RST: 1
Example: See Chapter 8.2.4, "Data handling", on page 1511.
Usage: Setting only
SCPI command reference