5.11.3.1 Controls on the Pwr Cal Settings tab
Some buttons serve as openers for additional dialogs:
"Cal Power...": see Chapter 5.11.3.2, "Modify Cal Power dialog", on page 527
"Transm. Coefficients...": see Chapter 5.11.3.3, "Power Meter Transmission Coeffi-
cients dialog", on page 529
"Power Meters...": see Chapter 4.7.27, "External power meters", on page 265
Switch Off Other Sources
Ensures that the power at all sources except the calibrated source is switched off dur-
ing the calibration. Recommended, if the measurement involves a combination of dif-
ferent signals.
Remote command:
SOURce<Ch>:POWer:CORRection:OSOurces[:STATe]
Flatness Cal – Max Iterations
Sets a limit for the number of calibration sweeps. See also "Calibration procedure"
on page 195.
Remote command:
SOURce:POWer:CORRection:NREadings
SOURce:POWer:CORRection:COLLect:AVERage[:COUNt]
Flatness Cal – Tolerance
Defines the maximum deviation of the measured power from the cal power. The cali-
bration procedure is stopped if "Max Iterations" is reached or if the measured power is
within the "Tolerance".