EasyManua.ls Logo

R&S ZNB Series - 8.2.4.4 Exporting S-parameters; 8.2.5 Calibration; 8.2.5.1 One and two-port calibration

R&S ZNB Series
1611 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Programming examples
R&S
®
ZNB/ZNBT
1515User Manual 1173.9163.02 ─ 62
8.2.4.4 Exporting S-parameters
The calibration defines which S-Parameters are allowed to be exported to a Touch-
stone file. In the following example, a default (TOSM) calibration is created to make all
S-Parameters available.
// Reset the analyzer
*RST
:SYSTEM:DISPLAY:UPDATE ON
:SENSe1:CORRection:COLLect:METHod:DEFine 'Test', TOSM, 1, 2, 3, 4
:SENSe1:CORRection:COLLect:SAVE:SELected:DEFault
// Initiate a complete sweep
:INITiate1:CONTinuous OFF; :INITiate:IMMediate;*WAI
:MMEMory:STORe:TRACe:PORTs 1, 'ParserTouchstonePorts.s1p', COMPlex, 2
:MMEMory:STORe:TRACe:PORTs 1, 'ParserTouchstonePorts.s2p', COMPlex, 3, 2
:MMEMory:STORe:TRACe:PORTs 1, 'ParserTouchstonePorts.s4p', COMPlex, 1, 4, 3, 2
8.2.5 Calibration
The following programming examples are related to system error correction and power
calibration.
8.2.5.1 One and two-port calibration
The following example calibrates one or two analyzer ports.
// Reset the analyzer
*RST
:SYSTEM:DISPLAY:UPDATE ON
// Set cal kit as active kit for N50
:SENSE:CORRECTION:CKIT:N50:SELECT 'ZV-Z121'
// Select connectors for the ports
:SENSE1:CORRECTION:COLLECT:CONNECTION1 N50MALE
:SENSE1:CORRECTION:COLLECT:CONNECTION2 N50MALE
// Don't save the cal standard measurements with apply cal, i.e. with the commands
// :SENSE1:CORRECTION:COLLECT:SAVE or
// :SENSe1:CORRection:COLLect:SAVE:SELected
// Instead, use the global, channel-independent setting:
:SENSe:CORRection:COLLect:ACQuire:RSAVe:DEFault OFF
Condensed programming examples

Table of Contents

Related product manuals