Command reference
R&S
®
ZNB/ZNBT
1065User Manual 1173.9163.02 ─ 62
Parameters:
<SmoothAperture> Smoothing aperture. A smoothing aperture of n % means that
the smoothing interval for each sweep point i with stimulus value
x
i
is equal to [x
i
- span*n/200, x
i
+ span*n/200], and that the
result of i is replaced by the arithmetic mean value of all mea-
surement points in this interval.
Range: 0.05% to 100%.
*RST: 1
Default unit: %
Example:
*RST; :CALC:SMO ON
Activate smoothing for the default trace.
CALC:SMO:APER 0.5
Reduce the smoothing aperture to 0.5 %.
Manual operation: See "Aperture" on page 362
CALCulate<Chn>:SMOothing[:STATe] <Boolean>
Enables or disables smoothing for trace no. <Chn>.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF - smoothing on or off.
*RST: OFF
Example: See CALCulate<Chn>:SMOothing:APERture
Manual operation: See "Smoothing" on page 362
CALCulate:TDVSwr[:STATe] <Boolean>
Enables/disables time domain site VSWR measurements.
Parameters:
<Boolean> ON | OFF – TD-VSWR on or off.
*RST: OFF
Manual operation: See "TD-VSWR" on page 350
7.3.2 CONFigure commands
The CONFigure... commands create and delete channels or traces, assign channel
and trace names, and define the channel sweep order. The commands are device-spe-
cific.
CONFigure:CHANnel:CATalog?.....................................................................................1066
CONFigure:CHANnel:MEASure:ALL[:STATe].................................................................. 1066
CONFigure:CHANnel:MEASure:OPTimized....................................................................1067
SCPI command reference

Table of Contents

Related product manuals