Parameter Meaning Comment/Unit
<Z0> Reference impedance (no unit) Must be specified without unit
(implicit unit is Ω)
<C0>,...,<C3> Polynomial coefficients for the fringing capacitance of the standard (load
parameter)
Must be specified without unit:
implicit unit of <Ci> is fF / (GHz)
i
<L0>,...,<L3> Polynomial coefficients for the residual inductance of the standard (load
parameter)
Must be specified without unit:
implicit unit of <Li> is pH / (GHz)
i
OPEN | SHORt |
MATCh |
<Resistance>
A load circuit model generally consists of a capacitance (modeled by
<C0>,...,<C3>), connected in parallel to an inductance (modeled by
<L0>,...,<L3>) and a resistance, both connected in series.
OPEN | SHORt | MATCh indicates a simplified modeling as an Open or
Short or Match standard.
OPEN: the resistance is infinite so that the standard behaves like a
capacitor (no inductance)
SHORt: the resistance is zero so that the standard behaves like an
inductance (no capacitance)
MATCh: the standard behaves like a match (no inductance, no
capacitance, resistance Z0)
<Resistance> indicates the general load circuit model.
Character data | numeric value
<Port1>,
<Port2>
Optional port restriction: one port number for one-port standards, two port
numbers for two-port standards
Integer values
The different standard types are defined by the following parameters. Port restrictions
are indicated in brackets:
Table 7-15: Standard types and their parameters
<std_type> Meaning
MOPen | FOPen Open: male (m) or female (f)
MSHort | FSHort Short: m or f
OSHort[<1|2|3>] | MOSHort[<1|2|
3>] | FOSHort[<1|2|3>]
Offset short: sexless, m or f (three standards each)
For user-defined connector types only.
Suffix 1 can be omitted.
MMTCh | FMTCh Match: m or f
MSMatch | FSMatch Sliding match: m or f
MREFlect | FREFlect Reflect: m or f
MMTHrough | MFTHrough |
FFTHrough
Through: m-m or m-f or f-f
MMLine[<1|2|3>] | MFLine[<1|2|3>]
| FFLine[<1|2|3>]
MMLine[<1|2|3>](P2P3) ...
Line: m-m or m-f or f-f (three standards each)
Suffix 1 can be omitted.
MMATten | MFATten | FFATten Attenuation: m-m or m-f or f-f
MMSNetwork | MFSNetwork |
FFSNetwork
Symmetric network: m-m or m-f or f-f
SCPI command reference