Command reference
R&S
®
ZNB/ZNBT
1238User Manual 1173.9163.02 ─ 62
Example: Suppose that the cal pool contains a file 'Calgroup3.cal'
with a valid through calibration for the active channel no. 1,
which you want to include in a new TOSM calibration for ports 1
and 2.
MMEM:CDIR DEF; CDIR 'Calibration\Data'
Go to the cal group directory.
:SENSe1:CORRection:COLLect:METHod:DEFine 'New
Cal', TOSM, 1, 2
Define a new TOSM calibration.
SENSe1:CORRection:COLLect:LOAD:SELected 'Cal
Group 3.cal', THROUGH, 1, 2
Load the through data from the cal pool file into the new calibra-
tion.
SENSe1:CORRection:COLLect:ACQuire:SELected
OPEN, 1
Proceed with the new calibration measurements.
Manual operation:
See "Start Cal Sweep" on page 470
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine <CalName>, <CalType>,
<TestPort1>[, <TestPort2>[, <TestPort3orAdapterCalKit>[, <TestPorts>]...]]
Defines the calibration to be performed for channel <Ch>.
If multiple channels shall be calibrated in parallel, use
●
[SENSe:]CORRection:COLLect:CHANnels:ALL ON or
●
[SENSe:]CORRection:COLLect:CHANnels:MCTYpes ON
before executing this command. For background information see Chapter 4.5.8, "Paral-
lel calibration of multiple channels", on page 201.
Calibration data is acquired using [SENSe<Ch>:]CORRection:COLLect[:
ACQuire]:SELected.
Suffix:
<Ch>
.
Channel number
Parameters:
<CalName> Name of the calibration (string parameter). The name serves as
a reference to delete a particular set of system correction data
([SENSe<Ch>:]CORRection:COLLect:DELete).
<CalType> ARTosm | ARPTosm | REFL | RSHort | FOPort | FRTRans |
FTRans | RTRans | OPTPort | POPTport | TOSM | TOM | TSM |
TRM | TRL | TNA | UOSM | PTOSm | PTOM | PTSM | PTRM |
PTRL | PTNA | PUOSm | MTOSm | MUOSm | MOSM | MTRL
Calibration types; see list below.