Command reference
R&S
®
ZNB/ZNBT
1166User Manual 1173.9163.02 ─ 62
MMEMory:STORe:CABLe <OutputDir>
Saves all predefined and user-defined Distance to Fault (DtF) cable types to the speci-
fied output directory.
Each DtF cable type is stored in a separate ASCII file <cable name>.rsc. The com-
mand silently overwrites files of the same name previously existing in the output direc-
tory.
Setting parameters:
<OutputDir> Output Directory, must be created before the command is exe-
cuted
Example: see MMEMory:LOAD:CABLe on page 1145
Usage: Setting only
Options: R&S ZNB/ZNBT-K3
Manual operation: See "Add / Delete" on page 733
MMEMory:STORe:CKIT
<CalKitName>, <CalKitFile>
Stores the data of a calibration kit to a specified file. The calibration kit is identified by
its name.
Setting parameters:
<CalKitName> Name of a user-defined calibration kit available on the analyzer.
Tip: It is not possible to modify or store predefined or ideal kits.
<CalKitFile> String parameter to specify the name and directory of the cal kit
file to be created. The file is a network analyzer-specific cal kit
file with the extension *.calkit.
If no path is specified the analyzer searches the current direc-
tory, to be queried with MMEMory:CDIRectory
Example:
MMEM:LOAD:CKIT 'C:\Users\Public\Documents
\Rohde-Schwarz\Vna\Calibration\Kits\New Kit
1.calkit'
Load the previously created cal kit file New Kit 1.calkit
from the default cal kit directory.
... :MMEM:STOR:CKIT 'New Kit 1', 'C:
\Users\Public\Documents
\Rohde-Schwarz\Vna\Calibration\Kits\New Kit
1.calkit'
Store the data for the user-defined cal kit "New Kit 1" and over-
write the cal kit file New Kit 1.calkit.
Usage:
Setting only
Manual operation: See "Import Cal Kit... / Export Cal Kit..." on page 516
SCPI command reference