Command reference
R&S
®
ZNB/ZNBT
1363User Manual 1173.9163.02 ─ 62
TSTamp
Timestamp in UTC
LTSTamp
Timestamp in local time
TVNA
For calibrations involving switch matrices, this indicates the test
ports on the VNA itself. Returns a comma-separated list of port
pairs TestPort,VnaPort
MVNA
For calibrations involving switch matrices, this indicates the
mapping between matrix ports and VNA ports for matrix
<Index>. Returns a comma-separated list of port pairs
MatrixVnaPort,VnaPort
MTESt
For calibrations involving switch matrices, this indicates the
mapping between matrix (physical) test ports and test ports for
matrix <Index>. Returns a comma-separated list of port pairs
MatrixTestPort,TestPort
FSMode
Returns the frequency sweep mode that was used during cali-
bration.
STEP: Stepped mode (for all segments)
ANAL: Swept mode (for at least one segment)
Because swept mode is limited to frequency sweeps this always
returns STEP.
<Index>
If one or more external switch matrices were used during cali-
bration, this refers to the index of a particular switch matrix (see
SYSTem:COMMunicate:RDEVice:SMATrix<Matr>:DEFine
on page 1405)
Usage: Query only
Manual operation: See "Port Overview" on page 478
SOURce<Ch>:POWer<PhyPt>:CORRection:DATA:PARameter<Wv>:COUNt?
Gets the number of available power calibrations (both source power and measurement
receiver).
Use SOURce<Ch>:POWer:CORRection:DATA:PARameter<Wv>? to retrieve the
related settings.
Suffix:
<Ch>
.
Number of the calibrated channel.
<PhyPt> This suffix is ignored.
<Wv> This suffix is ignored.
Usage: Query only
Manual operation: See "Port Overview" on page 478
SCPI command reference

Table of Contents